LXG CR

LXG CR

X-Ray Generator

Products

Shanghai Je-see Technology

   Application Scenarios

LXG CR 

Feature

  • Fully compliant with LXC testing

  • Integrated DAP function

  • Multi‑channel AEC function as standard

  • High compatibility, suitable for mainstream image software

  • Full‑digital stepless variable‑frequency rotary anode control technology

  • Modular and platform‑based design, easy for installation and maintenance

  • All‑in‑one aluminum oil tank with excellent sealing and pressure‑resistance

  • Digital adjustment of kV, mA and mAs

  • Dual‑filaments automatic selection


图片

Technical Data

Product Model             CR50CR80
Working Frequency≥400kHz
Input
Mains Phase/VoltageThree Phase/380VAC±10%
Power grid frequency50Hz/60Hz
Output
Nominal/Max Power50kW/50kW80kW/80kW
Tube Voltage40‑150 kV
Rising Time≤1.5ms@100KV[109%‑90%]
Accuracy of Tube Voltage±5%@≥100kV4kV@<100kV
Tube CurrentRAD:10‑710 mAContinuous Fluoro:05‑10mAPulsed Fluoro:0.5‑100 mARAD:10‑1000mAContinuous Fluoro:0.5‑10 mAPulsed Fluoro:0.5‑100 mA
Accuracy of Tube Current±(5%+1mA) @≥10ms±(20%+1mA) @<10ms
Current Time Product0.1‑1000 mAs
Accuracy of Current Time Product±(5%+0.2mAs)@≥10ms±(10%+0.2 mAs)@<10ms
Loading Time Range1‑10000ms
Accuracy of Loading Time±(2%+0.5ms)@≥10ms±(5%+1ms) @<10ms
Filament Frequency43kHz±2kHz
Rotor Control ModelFixed frequency/anode stepless speed control
Function/AccessoryAEC/DAP/Mini console/Touch screen console
Communication PortsRS232、CAN
LXCLXC testing report
Goods Size:W‑D‑H700mm400mm680mm

Real‑shot video