Technical Data
| Product Model | CRS50 | CRS80 | |
| Working Frequency | ≥400kHz | ||
| Input | |||
| Mains Phase/Voltage | Single Phase/220VAC±1096 | ||
| Power grid frequency | 50Hz/60Hz | ||
| Output | |||
| Nominal/Max Power | 32kW/32kW | 50kW/50kW | 80kW/80kW |
| Tube Voltage | RAD:40‑150 kV | ||
| Rising Time | <=1.5ms 100kv(10%‑90%) | ||
| Accuracy of Tube Voltage | ±56%≥100kV4kV<100kV | ||
| Tube Current | 10-400mA | 10-710mA | 10-1000mA |
| Accuracy of Tube Current | ±(50%+1mA) 0 <10ms | ||
| Current Time Product | 0.1-320mAs | 0.1-630mAs | 0.1-800mAs |
| Accuracy of Current Time Product | ±(5%+0.2mAs)≥10ms±(10%+0.2 mAs)<10ms | ||
| Loading Time Range | 1‑10000ms | ||
| Accuracy of Loading Time | ±(2%+0.5ms)≥10ms±(5%+1ms) <10ms | ||
| Filament Frequency | 43kHz±2kHz | ||
| Rotor Control Model | Fixed frequency/anode stepless speed control | ||
| Function/Accessory | AEC/DAP/Mini console/Touch screen console | ||
| Communication Ports | RS232、CAN | ||
| LXC | LXC testing report | ||
| Goods Size:L-W-H | 750mm * 470mm * 500mm | ||