LR01-XD60-1.2/5.5-40

LR01-XD60-1.2/5.5-40

Insert Tube

Products

Shanghai Je-see Technology

Application Scenarios

LR01-XD60-1.2/5.5-40

It is specially designed for mammography soft tissue radiography systems.

Technical Data

IEC ClassificationClass I Type B
Main Technical Indicators
Electrical ParameterCircuit

Power Source

(Neutral grounding)

Constant voltage type high

voltage generator

Nominal X-ray Tube VoltageWorking voltage22~40 kV
Nominal Focal Spot ValueLarge Focus.0.3 mm
Small Focus0.1 mm

Nominal Anode Input Power

(.1s)

Large Focus, 50HZ5.5 KW
Small Focus1.2 KW

 

Motor Rating
DutyDutyRunning
Power source (Hz)50/6050/60
Voltage (V)16060
Current (A)3.5~51.0~1.5
Min. Speed up (s)1~1.2---
Capacitor (μF)43/3043/30

 

Anode Speed
50 HZ2800 rpm
60 HZ3200 rpm
Stator impedance
Common Main Winding20.0Ω
common-auxiliary Winding50.0Ω

Resistance between Housing and Low

Voltage Terminals

2MΩMin

Normal operating range of the housing

temperature

16~75 ℃

 

Physical Parameter
DimensionsSee dimensional outline
Overall Length364.5 mm
Maximum Diameter138mm
Target
Angle12°
Diameter64mm
ConstructionMolybdenum Target
Total Filtration (Aluminum Equivalent)
Inherent Filtration1.0mmBe
Leakage Technique Factor40 kV、2.0 mA
X-ray Coverage276mm * 276mm at SID 650mm
WeightAbout 9.5 Kg
High Voltage terminalALDEN, P500
Cooling MethodNatural

 

Absolute Maximum and Minimum Ratings
(At any time, these values must not be exceeded.)
Maximum Tube Voltage40 kV
Minimum Tube Voltage22 kV
Filament Current
Small Focus4.5A
Large Focus7.1A
Filament Voltage
Small Focus3.5±1V
Large Focus12V
Filament Frequency Limits.0–25 kHz
Thermal Characteristics
Anode Heat Storage150kJ (200 kHU)
Maximum Anode Heat Dissipation300W (425 kHU)
X-ray Tube Assembly Heat Content750kJ (500kHU)
Normal Continuous Input Power
Without Air-circulator72W (6 kHU / min)

Tube Series

XD7-1.05XD6C-1.1XD4-2.9

XD3AXD2-1XC19

XD13CQ